Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Contrast-to-noise ratio difference in small field of view cone beam computed tomography machines
Boulos Bechara1, Clyde A McMahan, William S Moore
1Department of Oral and Maxillofacial Radiology, University of Texas Health Science Center at San Antonio, San Antonio, TX 78229-3900, USA. boulosbchara@hotmail.com
Abstract:
Each of three cone beam computed tomography (CBCT) machines offer a small field of view (FOV); Promax, Accuitomo 3D, and Kodak 9000. The objectives were to determine, when using a small FOV, the difference in contrast-to-noise ratios (CNR) of images obtained from different machines and different scan settings within each machine. A phantom made of a dry mandible with an epoxy resin based substitute (ERBS) block was used. The ERBS block was fixed to the lingual cortex in the posterior left mandible. Water was used to simulate soft tissue. Twenty scans were acquired. CNR was calculated as the difference in density of the ERBS block and a control area divided by the standard deviation of the difference. The CNR obtained from images acquired with the Promax machine (voxel size of 0.2 mm) was higher than the other machines (voxel sizes 0.076, 0.08, or 0.125 mm) when using a small FOV. The Accuitomo 360° scan had higher CNR than the Accuitomo 180° scan due to the doubling of the number of basis images acquired. Smaller voxel size does not necessarily lead to enhanced CNR. Increasing the number of basis images did increase the CNR.
More Related Videos
11:27Preparation and In Vitro Characterization of Dendrimer-based Contrast Agents for Magnetic Resonance Imaging
Published on: December 4, 2016
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...