Related Experiment Video
Updated: May 17, 2026

Quantitative Autonomic Testing
Published on: July 19, 2011
Automated vulnerability testing identifies patients with inadequate defibrillation safety margin
Ulrika Birgersdotter-Green1, Linda L Ruetz, Kishlay Anand
1University of California San Diego, San Diego, CA, USA. ubgreen@ucsd.edu
This study tested an automated system for implantable cardioverter-defibrillator testing, reducing the need for induced ventricular fibrillation (VF). The system successfully identified patients needing adjustments, proving feasible for safer, more efficient defibrillator implantation.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Medical Device Technology
Background:
- Implantable cardioverter-defibrillator (ICD) systems require efficacy testing at implantation, typically via induced ventricular fibrillation (VF).
- Assessing defibrillation safety margins without VF induction is possible using upper limits of vulnerability, but historically required manual T-wave timing.
- Manual T-wave timing is a limitation in current defibrillation safety margin assessments.
Purpose of the Study:
- To evaluate the feasibility of an inductionless system for implant testing of ICDs.
- To assess an automated vulnerability safety margin system that measures T-wave timing.
- To determine if automated testing can reduce the need for VF induction during ICD implantation.
Main Methods:
- A multicenter prospective study enrolled 60 patients undergoing ICD implantation.
- An automated system measured T-wave timing via intracardiac electrogram during ventricular pacing, delivering up to 4 T-wave shocks.
- Lack of VF induction by the automated system was considered adequate; patients then underwent conventional defibrillation testing.
Main Results:
- The automated system tested 54 patients, inducing VF in 10 (19%), with 2 requiring system revision.
- All patients not induced by the automated system were successfully defibrillated at ≤25 J.
- Bayesian statistics indicated a 97% to 100% credible interval for defibrillation success at 25 J in non-induced patients.
Conclusions:
- The automated system effectively identified patients failing conventional safety margin testing while minimizing VF induction.
- This automated approach shows feasibility for reducing the need for VF induction in ICD implant testing.
- Further studies with larger sample sizes are warranted to confirm these findings.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Automated Microbial Diagnostics
Dysrhythmias V: Evaluating Dysrhythmias
Errors occurring during blood pressure monitoring
Several factors...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation