Related Experiment Video
Updated: May 17, 2026

11:11
Identification of Novel CK2 Kinase Substrates Using a Versatile Biochemical Approach
Published on: February 21, 2019
Protein kinase inhibitor analysis with the core-shell separation technique.
Marek Dziadosz1, Rüdiger Lessig, Heidemarie Bartels
1Institute of Forensic Medicine, Otto-von-Guericke-University, Magdeburg, Germany. analytiker@chemist.com
Analytical and Bioanalytical Chemistry
|October 12, 2012
Summary
Core-shell separation technology enhances protein kinase inhibitor analysis using HPLC-DAD. This cost-effective method improves quantification limits, approaching LC-MS/MS capabilities for drug discovery.
Area of Science:
- Analytical Chemistry
- Chromatography
- Pharmaceutical Analysis
Background:
- Protein kinase inhibitors are crucial in cancer therapy.
- Accurate and sensitive analysis of these inhibitors is essential for drug development.
- Existing analytical methods may have limitations in sensitivity and efficiency.
Purpose of the Study:
- To evaluate the application of core-shell separation technology for protein kinase inhibitor analysis.
- To compare the performance of core-shell columns with conventional columns in HPLC-DAD.
- To assess the cost-effectiveness and sensitivity improvements offered by this technology.
Main Methods:
- Utilized a Kinetex 2.6 μm C8 100 Å core-shell HPLC column for separation.
- Employed a mobile phase of 0.05 M H(3)PO(4)/KH(2)PO(4) (pH = 2.3)-acetonitrile (7:3, v/v) at 0.7 mL/min.
- Performed sample preparation using liquid-liquid extraction with 1-chlorobutane.
Main Results:
- Achieved good linearity (5-10,000 ng/mL) and a limit of quantification (LOQ) of 5 ng/mL for canertinib.
- Core-shell columns produced 2-3 times narrower and higher peaks compared to conventional columns.
- Recoveries ranged from 74-79%, with relative standard deviation and accuracy below 15%.
Conclusions:
- Core-shell separation technology offers a cost-effective enhancement for HPLC-based protein kinase inhibitor analysis.
- This approach significantly improves peak characteristics and sensitivity, approaching LC-MS/MS performance.
- The validated method demonstrates suitability for the precise quantification of protein kinase inhibitors.