[Non-functional lead as a cause of detection disturbances in implantable cardioverter-defibrillator]
Maciej Kempa1, Szymon Budrejko, Mariusz Nowakowski
1Klinika Kardiologii i Elektroterapii, Gdański Uniwersytet Medyczny, 80–952 Gdańsk. kempa@gumed.edu.pl
Abstract:
The article presents the case of a patient treated with implantable cardioverter-defibrillator, in whom an additional lead (integrated bipolar) was implanted because of the damage to the ventricular lead. Interactions between the electrodes led to detection disturbances. The patient required the removal of the non-functional lead.