Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Robust algorithm for aligning two-dimensional chromatograms
Jonas Gros1, Deedar Nabi, Petros Dimitriou-Christidis
1Environmental Chemistry Modeling Laboratory (LMCE), GR C2 544, Swiss Federal Institute of Technology at Lausanne (EPFL), Station 2, CH-1015 Lausanne, Switzerland.
Abstract:
Comprehensive two-dimensional gas chromatography (GC × GC) chromatograms typically exhibit run-to-run retention time variability. Chromatogram alignment is often a desirable step prior to further analysis of the data, for example, in studies of environmental forensics or weathering of complex mixtures. We present a new algorithm for aligning whole GC × GC chromatograms. This technique is based on alignment points that have locations indicated by the user both in a target chromatogram and in a reference chromatogram. We applied the algorithm to two sets of samples. First, we aligned the chromatograms of twelve compositionally distinct oil spill samples, all analyzed using the same instrument parameters. Second, we applied the algorithm to two compositionally distinct wastewater extracts analyzed using two different instrument temperature programs, thus involving larger retention time shifts than the first sample set. For both sample sets, the new algorithm performed favorably compared to two other available alignment algorithms: that of Pierce, K. M.; Wood, Lianna F.; Wright, B. W.; Synovec, R. E. Anal. Chem.2005, 77, 7735-7743 and 2-D COW from Zhang, D.; Huang, X.; Regnier, F. E.; Zhang, M. Anal. Chem.2008, 80, 2664-2671. The new algorithm achieves the best matches of retention times for test analytes, avoids some artifacts which result from the other alignment algorithms, and incurs the least modification of quantitative signal information.
More Related Videos
08:232 in 1: One-step Affinity Purification for the Parallel Analysis of Protein-Protein and Protein-Metabolite Complexes
Published on: August 6, 2018
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Principles Of Column Chromatography
Chromatographic Methods: Terminology
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Affinity Chromatography