Introduction:
Thiopental is a cornerstone in the treatment of refractory status epilepticus and intractable intracranial hypertension. In our center we observed that thiopental might cause falsely elevated serum sodium levels.
Methods:
Triggered by a recent case experience of extremely elevated serum sodium levels during thiopental treatment, we retrospectively identified 53 patients treated with thiopental in our intensive care unit between 2007 and 2011 and evaluated electrolyte changes. We differentiated the analysis before and after introduction of a new device for sodium assays (Dimension Vista, Siemens) in the central laboratory in April 2010. Standardized in vitro laboratory tests were performed to study the effect of thiopental on sodium analysis.
Results:
Before April 2010, serum sodium levels determined in the central laboratory showed a good agreement with the bedside point-of-care (POC) device during thiopental therapy with [sodium](laboratory) - [sodium](POC) of only 1.08 mmol/L (P = .0517). After April 2010, a strong discrepancy between laboratory values and POC values was observed with [sodium](laboratory) - [sodium](POC) = 11.57 mmol/L (P < .0001). Standardized in vitro testing confirmed that thiopental induced a dose-dependent false hypernatremia (P = .002).
Conclusions:
Thiopental treatment can result in falsely elevated serum sodium. This is a critical finding since high sodium levels preclude administrating mannitol or hypertonic saline for the treatment of elevated intracranial pressure. Moreover, a false high sodium level might lead to the inappropriate administration of hypotonic fluids potentially resulting in increased brain edema and even higher intracranial pressure. To our knowledge, this is the first paper describing this clinically relevant phenomenon.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Ionic Bonds
When atoms gain or lose electrons to achieve a more stable electron configuration they form ions. Ionic bonds are electrostatic attractions between ions with opposite charges. Ionic compounds are rigid and brittle when solid and may dissociate into their constituent ions in water. Covalent compounds, by contrast, remain intact unless a chemical reaction breaks them.
Opposing Charges Hold Ions Together in Ionic Compounds
Ionic bonds are reversible electrostatic interactions between ions...
Parenteral Anesthetics: Overview
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Cardiac Catheterization I: Pre-Procedure Overview