Which point-of-care creatinine analyser for radiology: direct comparison of the i-Stat and StatStrip creatinine
Goce Dimeski1, Vera Tilley, Brock W Jones
1Pathology Queensland, Chemical Pathology, Princess Alexandra Hospital, Ipswich Road, Woolloongabba, Brisbane, QLD 4102, Australia. goce_dimeski@health.qld.gov.au
Background:
Availability of whole blood creatinine estimation for patients scheduled to undergo radiological contrast investigations can provide information to aid patient care by reducing adverse effects and improving departmental efficiencies.
Methods:
We performed imprecision studies, different patient sample type comparison in 40 participants, and a limited interference study with dopamine and dobutamine on the i-Stat and StatStrip point-of-care enzymatic analysers with the Beckman DxC800 Jaffe assay.
Results:
Imprecision results showed that the i-Stat performed better. Patient comparison data indicated that the i-Stat provided better correlation than the StatStrip for all the different sample types with correlation coefficients (r(2)) being 0.995-0.996 and 0.918-0.995, respectively. The i-Stat results had a small positive bias of 6-9% for the three different sample types, which required different reference intervals. The StatStrip method showed greater scatter and overall small negative bias of -6% for the whole blood samples and a 10% positive bias with the plasma samples. Dopamine caused significant positive interference with the i-Stat only while dobutamine caused a small negative bias with the StatStrip method only.
Conclusions:
The findings indicated there are differences offered by the two systems. The StatStrip requires a very small finger prick capillary sample, calculates estimation of the glomerular filtration rate and has an adjustment option to improve correlation with the local method. The i-Stat offers better analytical imprecision and patient comparison with the laboratory method with the three sample types but showed significant interference from dopamine. A final consideration was the availability of middleware to capture patient results with the i-Stat. Based on all the study data, the i-Stat was recommended.
Related Concept Videos
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Automated Microbial Diagnostics