Performance of an automatic dose control system for CT: patient studies
1Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Leipzig.
Purpose:
To study the effect of an automatic dose control (ADC) system with adequate noise characteristic on the individual perception of image noise and diagnostic acceptance compared to objectively measured image noise and the dose reductions achieved in a representative group of patients.
Materials And Methods:
In a retrospective study two matched cohorts of 20 patients each were identified: a manual cohort with exposure settings according to body size (small - regular - large) and an ADC cohort with exposure settings calculated by the ADC system (DoseRight 2.0™, Philips Healthcare). For each patient, 12 images from 6 defined anatomic levels from contrast-enhanced scans of chest and abdomen/pelvis were analyzed by 4 independent readers concerning image noise and diagnostic acceptance on a five-point Likert scale and evaluated for objectively measured image noise. Radiation exposure was calculated from recorded exposure data.
Results:
Use of the ADC system reduced the average effective dose for patients by 36 % in chest scans (3.2 vs. 4.9 mSv) and by 17 % in abdomen/pelvis scans (7.6 vs. 8.3 mSv). Average objective noise was slightly lower in the manual cohort (11.1 vs. 12.8 HU), correlating with a slightly better rating in subjective noise score (4.4 vs. 4.2). However, diagnostic acceptance was rated almost equal in both cohorts with excellent image quality (4.6 vs. 4.5).
Conclusion:
Use of an ADC system with adequate noise characteristic leads to significant reductions in radiation exposure for patients while maintaining excellent image quality.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...