Related Experiment Video
Updated: May 17, 2026

Full-Field Optical Coherence Microscopy for Histology-Like Analysis of Stromal Features in Corneal Grafts
Published on: October 21, 2022
Depth of intrastromal corneal ring segments by OCT
Modi Naftali1, Haneen Jabaly-Habib
1Ophthalmology Unit of the Baruch Padeh Medical Center, Poriya, Israel; Faculty of Medicine, Bar Ilan University, Galille, Israel. mnaftali@poria.health.gov.il
Purpose:
To compare the depth of intrastromal corneal ring segments (ICRS) with the expected depth value using optical coherence tomography (OCT).
Methods:
This was a retrospective comparative study in an ophthalmic unit in a government hospital, the Baruch Padeh Medical Center, Poriya, Israel. Ten eyes of 8 patients with 18 ICRS were reviewed. Eleven segments were Intacs (Addition Technology, Inc.) and 7 Kerarings (Mediphacos). Using anterior segment OCT (OPKO OTI) the shortest distance from the epithelium to the segment at 3 points was measured for each segment. The 3 points are proximal, middle, and distal to the incision.
Results:
The mean depth of the 18 segments was 360±68 µm. The mean maximal and minimal depths were 383±70 and 336±72 µm, respectively. The mean depths of the distal, central, and proximal point measurements of all ICRS were 358±79, 361±77, and 362±59 µm, respectively; no significant difference was found. No part of the segments tended to be more superficial than others (p=0.98). There was no significant difference between Intacs and Kerarings depths (p=0.43). There was a significant difference between the expected ICRS depth and the OCT measurements (mean 480±20) and 360±68), respectively.
Conclusions:
The ICRS actual depth was less than expected. There was mild variability in segment depth, both between segments and along the same segment. No significant difference was found between the depth of Intacs and Kerarings.