Effect of serum lyophilization on the rate constants of enzymatic methods for measuring cholesterol
1Clinical Pathology Department, Warren G. Magnuson Clinical Center, National Institutes of Health, Bethesda, MD 20892.
Abstract:
We determined the equilibrium absorbances and rate constants for two enzymatic methods, aca (DuPont) and RA-1000 (Technicon), used in determining cholesterol in reconstituted lyophilized serum. The lyophilized materials included two serum pools, three control materials, a College of American Pathologists' survey material, and Standard Reference Material no. 909. We calibrated the reagents with aca standards for cholesterol (DuPont). The difference in the mean concentrations of cholesterol (aca - RA-1000) was -0.09 g/L overall and was not statistically significant by analysis of variance. The mean rate constant for all materials was 0.23 min-1 for the aca and 1.42 min-1 for the RA-1000, significantly different (P less than 0.001). Lyophilization causes lower results for the aca method than for the Ra-1000, because the reaction rate for the aca method is slower and has not reached equilibrium when the final absorbance reading is made.
Related Concept Videos
Enzyme Kinetics
Scientists typically study enzyme kinetics with a fixed amount of enzyme in the controlled environment of a test tube. When more reactant, or substrate, is...
Measuring Reaction Rates
Introduction to Enzyme Kinetics
The experimenter can then plot the initial reaction rate or velocity (Vo) of a given trial against the substrate concentration ([S]) to obtain a graph of the reaction properties. For many enzymatic reactions involving a...