Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Development and characterization of a field-deployable ion-trap mass spectrometer with an atmospheric pressure
Alexander Misharin1, Konstantin Novoselov, Victor Laiko
1MassTech, Inc., 6992 Columbia Gateway Drive, Columbia, Maryland 21046, United States. alexander.misharin@apmaldi.com
Abstract:
A field-deployable quadrupole ion-trap mass spectrometer with an atmospheric pressure interface is designed, built, and characterized. The instrument enclosure (48 cm × 43 cm × 42 cm) includes a roughing pump and a helium lecture bottle; the total weight of the instrument is 68 kg. Peak power consumption during the instrument operation is ∼500 W. The instrument has a mass range of m/z 30-2500, across which it provides better than unit mass resolution. The typical peak width at half height is 0.3 Th for a scan speed of 4000 Th/s. Operation of the instrument with electrospray and atmospheric-pressure matrix-assisted laser desorption ionization (AP-MALDI) ion sources is demonstrated. AP-MALDI analysis of low femtomole amounts of peptides reveals that the sensitivity of the instrument is on par with modern commercially available quadrupole ion-trap mass spectrometers. Tandem mass spectrometry capabilities of the instrument include simultaneous isolation and fragmentation of several different compounds. Two ways to reduce the size, weight, and power consumption of the portable instrument were explored, and results of these initial studies are presented. One of the ways includes utilization of hydrogen as a buffer gas for operation of the ion-trap mass analyzer in combination with a metal hydride method for storage of hydrogen in a compact rechargeable cartridge. Furthermore, careful selection of the inlet capillary dimensions allowed to eliminate the first "1 Torr" stage of the differential pumping without any significant loss of the instrument sensitivity. The elimination of this first pumping stage removed two turbo drag pumps, which substantially decreased the instrument's maximum power consumption (to ∼300 W in peak use, and ∼150 W during standby), as well as its size (to 30 cm × 43 cm × 50 cm) and weight (to 35 kg).
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometers
Mass Spectrum: Interpretation
Tandem Mass Spectrometry
Mass Analyzers: Overview