Predictors of high defibrillation threshold in the modern era
Edward F Lin1, Darshan Dalal, Alan Cheng
1Johns Hopkins University, Baltimore, Maryland, USA.
Introduction:
High defibrillation threshold (DFT) is a clinical problem in 1-8% of implantable cardioverter-defibrillator implants. Some clinicians and investigators question whether the benefits of routine DFT testing outweigh the risks. Identification of the predictors of elevated DFT may allow selective application of DFT testing. However, the clinical characteristics of patients with high DFT in the modern era have not been well-defined.
Methods:
All patients who underwent DFT testing in our institution during an 8-year period were reviewed for this retrospective study. High DFT was defined as less than a 10-J safety margin on initial testing. For each case, the two cases preceding and two cases following by the same implanter were selected as controls.
Results:
Of the 2,138 patients who underwent DFT testing, 48 (2.2%) met criteria for high DFT. Compared to 192 control patients, patients with high DFT were more likely to be younger (P = 0.004), have nonischemic cardiomyopathy (P = 0.036), have a longer QRS interval (P = 0.026), and have a left ventricular ejection fraction (LVEF) ≤ 0.25 (P = 0.013). On multivariate analysis, only younger age (P = 0.016) and LVEF ≤ 0.25 (P = 0.010) remained statistically significant predictors of elevated DFT.
Conclusions:
High DFT was identified in 2.2% of ICD implants in our institution in recent years. Although younger age and depressed LVEF predicts this problem, elevated DFT occurred in patients of all ages and ejection fractions. Elimination of routine DFT testing appears to be premature given the prevalence and unpredictability of elevated DFT.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...