Related Experiment Video
Updated: May 17, 2026

U-Shaped Horizontal Swimming Technique for Preparing High-Quality Sperm with Low DNA Fragmentation Index
Published on: March 28, 2025
A microfluidic device to reduce treatment time of intracytoplasmic sperm injection
Koji Matsuura1, Takuya Uozumi, Takuya Furuichi
1Research Core for Interdisciplinary Sciences, Okayama University, Okayama, Japan. kojimatu@md.okayama-u.ac.jp
Objective:
To develop a microfluidic device that can reduce the intracytoplasmic sperm injection (ICSI) treatment time by increasing sperm concentration.
Design:
We compared the ICSI treatment time required for porcine sperm using a method employing the microfluidic device and one using the conventional microdroplet method.
Settings:
Academic research laboratories at Okayama University.
Animal(S):
Reproductive cells of porcine sperm, oocytes, and embryos.
Intervention(S):
Cell manipulations, ICSI, and embryo culture.
Main Outcome Measure(S):
Average ICSI treatment time and sperm concentration.
Result(S):
The average ICSI treatment time (mean ± SEM) using the method with the microfluidic device for poor-quality semen (sperm concentration, 2.0 × 10(4) cells/mL) was significantly shorter than the treatment time using the conventional microdroplet method (265 ± 15 seconds [n = 43] vs. 347 ± 19 seconds [n = 50]). When diluted semen with a sperm concentration of 2.0 × 10(5) cells/mL was used, no significant difference was observed between the two methods (n = 50 and n = 48).
Conclusion(S):
The microfluidic device can reduce the time required for ICSI treatment that is used to increase sperm concentration in poor-quality semen samples. The results suggest that this device may be clinically useful for ICSI treatment in human assisted reproductive technology.