Published on: June 16, 2022
Calcium-phosphate microprecipitates mimic microparticles when examined with flow cytometry
Michael C Larson1, Maia R Luthi, Neil Hogg
1Department of Biophysics, Medical College of Wisconsin, Milwaukee, Wisconsin 53226, USA. mlarson@mcw.edu
Summary
Calcium-phosphate microprecipitates can mimic microparticle signals in flow cytometry. Careful buffer selection is crucial to avoid false positives when detecting microparticles using annexin V, which requires calcium ions.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Hematology
Background:
- Circulating microparticles (MPs) are elevated in various disease states.
- Flow cytometry is a key technique for MP analysis, but distinguishing MPs from artifacts is challenging.
- Annexin V, a phosphatidylserine-binding protein, is commonly used to identify MPs, requiring millimolar calcium ion concentrations.
Purpose of the Study:
- To investigate the potential for calcium-phosphate microprecipitates to interfere with microparticle detection via flow cytometry.
- To assess the impact of calcium concentration on the flow cytometry signal of calcium-phosphate microprecipitates.
- To provide recommendations for avoiding artifacts in MP analysis.
Main Methods:
- Calcium-phosphate microprecipitates were generated by titrating calcium ions into phosphate-buffered saline (PBS).
- Microprecipitates and microparticles from aged donor blood were analyzed using flow cytometry.
- The effect of calcium chelators and fluorescently labeled antibodies on microprecipitate signals was examined.
Main Results:
- Calcium-phosphate microprecipitates exhibited flow cytometry signals overlapping with those of microparticles.
- Microprecipitate signals increased with calcium ion concentration and were abolished by calcium chelation.
- Fluorescent signals in biological samples showed an apparent increase with calcium, returning to baseline after chelation.
Conclusions:
- Calcium-phosphate microprecipitates can generate artifacts that mimic microparticle signals in flow cytometry.
- Avoiding phosphate-containing buffers or ensuring adequate calcium chelation is essential when using annexin V for MP detection.
- Flow cytometry may be useful for analyzing crystalline structures in biological fluids.