Related Experiment Video
Updated: May 17, 2026

Implementation of a Reference Interferometer for Nanodetection
Published on: April 26, 2014
Simultaneous measurement of external refractive index and temperature based on long-period-grating-inscribed Sagnac
Peng Gao1, Xiuyan Chen, Wenjiang Feng
1College of Physics Science and Technology, Shenyang Normal University, Shenyang 110034, People's Republic of China. gaopengsysd@126.com
Abstract:
A compact Sagnac interferometer with a long-period grating (LPG) inscribed in a polarization-maintaining fiber (PMF-LPG) and a fiber Bragg grating (FBG) is proposed. The PMF-LPG is intrinsically sensitive to the external refractive index (ERI) and temperature, whereas the FBG is only sensitive to temperature. Simultaneous measurement of the ERI and temperature can be achieved by detecting wavelength shifts in the interference patterns of the Sagnac interferometer and the FBG. Because a Sagnac interference structure is used, a high ERI sensitivity of -21.07 nm∕refractive index unit in measurement range 1.33-1.43 is obtained, which is about three times higher than for the normal LPG ERI sensor.
Related Concept Videos
IR Spectrometers
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...