Porphyrin-Modified Beads for Use as Compensation Controls in Flow Cytometry
Published on: March 24, 2023
Quality assurance for polychromatic flow cytometry using a suite of calibration beads
Stephen P Perfetto1, David Ambrozak, Richard Nguyen
1Flow Cytometry Core Facility, Vaccine Research Center, National Institute of Allergy and Infectious Diseases (NIAID), National Institutes of Health (NIH), Bethesda, MD, USA. sperfetto@mail.nih.gov
None:
The quality assurance program presented here provides a means to maximize and maintain the performance of individual flow cytometers in a facility. To optimize performance, we recommend performing all three steps (optimization, calibration and standardization) in this program when a new flow cytometer is installed or whenever the flow cytometer's optical path is altered (e.g., lasers, filters or detectors are replaced). The complete process requires 3-4 h. On a more frequent basis, only a subset of these procedures need to be performed as a part of daily maintenance routines. The data generated can be tracked to monitor the instrument and determine whether service is needed. In addition, the data can provide a metric for whether repairs and upgrades have improved or harmed performance, and for future instrument-to-instrument comparisons. In sum, the procedures presented here represent an updated framework for optimizing, calibrating and standardizing a flow cytometer for daily use.