Visceral adipose tissue quantification using Lunar Prodigy.

David L Ergun1, Megan P Rothney, Mary K Oates

  • 1GE Healthcare, Madison, WI, USA. ergundavid@gmail.com

Journal of Clinical Densitometry : the Official Journal of the International Society for Clinical Densitometry
|November 15, 2012
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Total-body versus knee-specific dual-energy X-ray absorptiometry scans for assessing bone mineral density surrounding the knee joint in collegiate athletes with and without an anterior cruciate ligament reconstruction.

Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry·2025
Same author

Diagnosing Osteoporosis for the Spine Practitioner.

Spine·2025
Same author

Trabecular bone score predicts distal femur screw pull-out strength better than DXA bone mineral density, CT Hounsfield units, and cortical density.

Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA·2025
Same author

Radiation exposure during simulated equine head and limb fan beam standing computed tomography appears safe for personnel using lead shielding.

Journal of the American Veterinary Medical Association·2024
Same author

Comparison of radiography and computed tomography for identification of third metacarpal structural change and associated assessment of condylar stress fracture risk in Thoroughbred racehorses.

Equine veterinary journal·2024
Same author

Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.

Endocrine reviews·2024

A new dual-energy X-ray absorptiometry (DXA) application, CoreScan, shows similar visceral adipose tissue (VAT) measurements between GE Prodigy and iDXA densitometers. However, the Prodigy exhibits lower precision, potentially requiring larger VAT changes for detection.

Area of Science:

  • Medical Imaging
  • Body Composition Analysis
  • Radiology

Background:

  • Dual-energy X-ray absorptiometry (DXA) is increasingly used for body composition analysis.
  • A novel DXA application, CoreScan, enables visceral adipose tissue (VAT) measurement in the android region.
  • Previous validation of CoreScan was performed on the Lunar iDXA densitometer.

Purpose of the Study:

  • To evaluate the agreement of CoreScan VAT measurements between GE Prodigy and Lunar iDXA densitometers.
  • To characterize differences in VAT measurement precision between the Prodigy and iDXA instruments.

Main Methods:

  • Paired DXA measurements using both Prodigy and iDXA densitometers were obtained from volunteers.
  • Empirical adjustments to VAT algorithm parameters were defined using a subset of data (n=59).

More Related Videos

Adipo-Clear: A Tissue Clearing Method for Three-Dimensional Imaging of Adipose Tissue
08:52

Adipo-Clear: A Tissue Clearing Method for Three-Dimensional Imaging of Adipose Tissue

Published on: July 28, 2018

Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging
08:31

Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging

Published on: July 1, 2021

Related Experiment Videos

Last Updated: May 16, 2026

Localization, Identification, and Excision of Murine Adipose Depots
08:53

Localization, Identification, and Excision of Murine Adipose Depots

Published on: December 4, 2014

Adipo-Clear: A Tissue Clearing Method for Three-Dimensional Imaging of Adipose Tissue
08:52

Adipo-Clear: A Tissue Clearing Method for Three-Dimensional Imaging of Adipose Tissue

Published on: July 28, 2018

Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging
08:31

Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging

Published on: July 1, 2021

  • Performance validation was conducted on a separate dataset (n=62).
  • Main Results:

    • Prodigy and iDXA CoreScan VAT measurements demonstrated high correlation (r=0.984).
    • The mean difference in VAT volume was -13.8 cm³ (95% CI: -45 to +17 cm³).
    • Short-term precision analysis revealed smaller measurement error variance for iDXA (7.3%) compared to Prodigy (9.8%).

    Conclusions:

    • CoreScan VAT measurements are comparable between GE Prodigy and Lunar iDXA densitometers.
    • The GE Prodigy densitometer exhibits lower VAT measurement precision than the Lunar iDXA.
    • Investigators using the Prodigy may need to target larger changes in VAT for reliable detection.