Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Atypical upswing at initial part of phase III in capnograms due to technical cause
1Department of Anesthesia and Pain Medicine, The Catholic University of Korea College of Medicine, Seoul, Korea.
Abstract:
We sporadically experienced three paediatric cases of atypical upswing at the initial part of phase III in the capnograms via side-stream capnometer immediately following endotracheal intubation and mechanical ventilation. No fault was found in the monitor or anaesthetic system including breathing circuits, carbon dioxide (CO2 ) sampling tube, water trap, and unidirectional valves. The upsurge of CO2 disappeared with increasing the respiratory rate; however, it reappeared with decreasing the respiratory rate and vice versa. We experimentally reproduced the phenomenon of overshooting CO2 measurement after the luer lock connection of the sample gas line at the water trap had been unscrewed a little bit.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Acute Respiratory Failure-III
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Atelectasis II: Pathophysiology
Respiratory Volumes and Capacities I