Ultrasonographic Evaluation of Breast Cancer-related Lymphedema
Published on: January 12, 2017
Use of impedance ratios to assess hand swelling in lymphoedema
E S Dylke1, H Alsobayel, L C Ward
1Faculty of Health Sciences, University of Sydney, Sydney, Australia.
Objectives:
To determine whether bioimpedance spectroscopy was suitable for detection of hand lymphoedema.
Methods:
The hands of 50 participants without a history of lymphoedema were measured with perometry and bioimpedance spectroscopy after positioning two ways for three minutes: (a) both hands rested at heart height and (b) the dominant hand at heart height and the non-dominant hand at head height. In addition, 10 women with secondary hand lymphoedema were also measured.
Results:
Impedance and volume measurements were found to be strongly related (dominant hand r = -0.794). Both measurements were reliable (ICC2,1 = 0.900-0.967 and 0.988-0.996, respectively). Impedance was more sensitive to small changes in hand volume due to the postural change (position × device interaction: F = 23.9, P < 0.001). Finally, impedance measurements had better discrimination of women with lymphoedema than volume measurements.
Conclusions:
Bioimpedance spectroscopy is a promising tool for the detection of secondary hand lymphoedema.
Related Concept Videos
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of the Cardiovascular System II: Inspection
Head and Neck