Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Evaluation of LOCI technology-based thyroid blood tests on the Dimension Vista analyzer
D Monneret1, D Guergour, S Vergnaud
1Department of Biochemistry Toxicology Pharmacology, University Hospital, Grenoble, France. dmonneret2@gmail.com
Objectives:
To assess the performance of thyroid-stimulating hormone (TSH), free thyroxine (FT4) and free triiodothyronine (FT3) determinations by luminescent oxygen channeling immunoassay (LOCI) technology on the Dimension Vista analyzer (Siemens Healthcare Diagnostics).
Design & Methods:
We assessed 1) functional sensitivity for TSH (FSe-TSH), and intra- and inter-assay variations for TSH, FT4 and FT3 on Vista; 2) comparisons of serum and heparin-treated plasma on Vista; 3) comparisons of a) plasma TSH by Vista versus electrochemiluminescence (ECLIA) on Roche Modular analyzer, and b) plasma FT4 and FT3 by Vista versus Immunotech-Beckman radioimmunoassay (RIA); and 4) association of albumin and prealbumin levels with free thyroid hormone concentrations on Vista.
Results:
1) FSe-TSH concentration was below 0.005 mIU/L. Maximum intra-assay CVs (2.1%, 1.4%, 5.2%) and inter-assay CVs (16.5%, 5.1%, 5.8%) were good for TSH, FT4 and FT3 respectively. 2) Heparin-treated plasma samples consistently gave slightly higher values than serum for TSH, FT4 and FT3. 3) Passing-Bablok regression gave: TSH: [LOCI]=0.91[ECLIA]-0.08 (concordance correlation coefficient ρ(c)=0.95); FT4: [LOCI]=1.05[RIA]-1.55 (ρ(c)=0.80); and FT3: [LOCI]=1.05[RIA]-0.06 (ρ(c)=0.81). 4) Both serum albumin and prealbumin concentrations were positively associated with FT3 levels and negatively associated with FT4 levels in patients.
Conclusion:
LOCI is accurate for TSH, FT4 and FT3 analysis. Despite a slight significant bias compared to ECLIA, LOCI is precise for TSH and fulfills the third-generation criteria. However, the poor concordance between LOCI and RIA for FT4 and FT3, and the dependence of these hormones on binding proteins require further investigation.
More Related Videos
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
05:41Synchronous Triplanar Reconstruction Integrated with Color Doppler Mapping for Precise and Rapid Localization of Thyroid Lesions
Published on: February 9, 2024
Related Concept Videos
Automated Microbial Diagnostics
Therapeutic Drug Monitoring: Drug Analysis Methods
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: