Validation of a High-Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment
Related Experiment Video
Updated: Nov 1, 2026

Capillary Electrophoresis Separation of Monoclonal Antibody Isoforms Using a Neutral Capillary
Published on: January 16, 2017
Kyle D Luttgeharm1, Manu Grover2, Sheng-Yuan Huang1
1Agilent Technologies, Ankeny, Iowa, USA.
Abstract:
Fluorescent capillary gel electrophoresis (CGE) with sodium dodecyl sulfate (CE-SDS) provides a powerful, high-sensitivity alternative to ultraviolet (UV)-based detection for characterizing therapeutic monoclonal antibodies (mAbs). Regulatory and standards organizations, such as the United States Pharmacopeia (USP), include only UV-based CE-SDS methods, hindering the adoption of alternative detection methods. There is growing opportunity to expand beyond exclusively UV-based CE-SDS methods. In this study, we present an analytical validation of a light-emitting diode (LED) fluorescence-based parallel CE-SDS method for both non-reduced and reduced analysis of therapeutic antibodies. Using the NISTmAb reference material as a model system, size heterogeneity critical quality attributes (CQAs), including monomeric purity (non-reduced conditions), percent glycosylation (reduced conditions), and percent thioether (reduced conditions), were assessed. The fluorescence method demonstrated high specificity and precision with relative standard deviation (RSD) values <1% for monomeric purity and glycosylation, and <3% for thioether, as well as robust performance across variations in injection voltage, electrophoresis voltage, labeling temperature, and Labeling Buffer concentration. Ruggedness testing across users and reagent lots confirmed reproducibility, and accuracy assessments showed strong agreement with reported values from the National Institute of Standards and Technology (NIST) and traditional UV detection measurements. Linearity studies yielded the coefficient of determination (R2) values >0.99 for both non-reduced and reduced analyses. These results highlight the high sensitivity, stable baseline performance, and suitability of LED fluorescence-based parallel CE-SDS as a validated, higher-throughput alternative to traditional UV-based methods for mAb quality control (QC).