Reliability and accuracy of a method of adjustment for self-measurement of auditory thresholds
Dianne J Van Tasell1, Paula Folkeard
1IntriCon Corporation, Arden Hills, Minnesota, USA.
Objective:
To evaluate the reliability and accuracy of a method for measuring pure tone air conduction thresholds in which the user adjusts test tones to threshold, using an iPad, automated instructions, and minimal supervision.
Study Design:
Prospective nonrandomized validation study.
Setting:
University hearing research laboratories and audiology clinics.
Patients:
Fifty-five adults with hearing loss in at least 1 ear ranging from mild to severe.
Intervention:
Automated measurement of pure tone air conduction thresholds using the following: a software-controlled adaptive method, and a user-controlled method of adjustment, both implemented on a calibrated iPad and using standard audiometry earphones.
Main Outcome Measure:
Test-retest reliability of both methods, comparison of thresholds measured with automated techniques to thresholds measured using manual audiometry.
Results:
For both iPad methods, test-retest differences were smaller than those reported in other studies for manual audiometry. Average automated versus manual threshold differences were within the range of expected variance of manual audiometry. Subjects preferred the adjustment method.
Conclusion:
Both iPad self-test methods yield accurate and reliable pure tone air conduction thresholds.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5% chance...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...