[Determination of nine kinds of polycyclic aromatic hydrocarbons in serum by high performance liquid chromatography]
Changming Ding1, Lei Zheng, Shaobin Lin
1Institute for Environment Hygiene and Health Related Product Safety, Chinese CDC, Beijing 100021, China.
Objective:
To develop a high performance liquid chromatography (LC) method for simultaneously determination of polycyclic aromatic hydrocarbons (PAHs) in serum.
Methods:
PAHs could be separated completely by using C18 column and detected by fluorescence detector at wavelengths (Ex = 340 nm, Em = 425 nm). The mobile phase A was acetonitrile and mobile phase B was water with a flow rate at 1.0 ml/min.
Results:
The concentration of PAHs achieved a good linear relation with the peak area. The correlation coefficient was in the range of 0.9941-0.9999. The recoveries were in the range of 91.4%-99.8% and the RSD were in the range of 2.56%-12.7%. The detection limits were in the range of 0.05-0.10 ng/ml.
Conclusion:
The method was simple, rapid, sensitive and accurate, and suitable for the detection of PAHs in serum.
More Related Videos
08:12Characterization, Quantification and Compound-specific Isotopic Analysis of Pyrogenic Carbon Using Benzene Polycarboxylic Acids (BPCA)
Published on: May 16, 2016
04:39A Modified QuEChERS-HPLC Method for Detection of Polycyclic Aromatic Hydrocarbons in Zebrafish Embryos Exposed to Fine Particulate Matter
Published on: June 13, 2025
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Instrumentation
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...