Optimization for Axial Resolution, Depth Range, and Sensitivity of Spectral Domain Optical Coherence Tomography at
Sang-Won Lee1, Hyun-Woo Jeong, Beop-Min Kim
1Research Institute of Health Sciences, Korea University, Seoul 136-703 and Department of Biomedical Engineering, Korea University, Seoul 136-703.
Abstract:
We have developed high-speed and high-resolution spectral domain optical coherence tomography at 1.3 µm using an InGaAs line-scan camera and a broadband light source with the bandwidth of 170 nm that produces a theoretical axial resolution of 4.4 µm in air. We compared axial resolutions from point spread functions (PSFs) and depth ranges while changing the full spectral bandwidth detected by the camera and describing the optimization process for the axial resolution, the depth range, and the sensitivity for SD-OCT system. We found that SD-OCT at 1.3 µm cannot satisfy the conditions both below the axial resolution of 5 µm and above the depth range of 2 mm because of the restricted pixel number of the line-scan camera. To scan a large depth range, the axial resolution has to be sacrificed. In addition, the sensitivity rolls off slowly as a function of the depth if a large depth range is scanned. On the other hand, if the axial resolution needs to be close to the theoretical one, the depth range becomes limited and the sensitivity decays quickly. Since we have to maintain a reasonable depth range of 2.0 mm, we chose the spectrum full bandwidth of 214 nm captured by the detector to balance the axial resolution of 8.2 µm. In this setting, the sensitivity of our OCT system was measured at 107.1 dB. Theoretical and experimental results are compared and presented in this paper.
More Related Videos
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...