Breast dynamic contrast-enhanced examinations with fat suppression: are contrast-agent uptake curves affected by
Maria A Schmidt1, M Borri, E Scurr
1Royal Marsden NHS Foundation Trust and Institute of Cancer Research, MRI Unit, Downs Rd, Sutton SM2 5 PT, UK. maria.schmidt@icr.ac.uk
Objectives:
To investigate the effect of magnetic field heterogeneity in breast dynamic contrast-enhanced examinations with fat saturation (DCE-FS).
Methods:
The magnetic field was mapped over the breasts in ten patients. DCE-FS was undertaken at 1.5 T with fast spoiled gradient echoes and spectrally selective fat saturation. Signal intensity was calculated for T1 values 25-1,200 ms both on and off resonance, and results were verified with a test object. Clinical examinations were evaluated for the predicted effects of field heterogeneity.
Results:
Magnetic field was found to vary by 3.6 ± 1.2 ppm over the central transaxial slice and 5.1 ± 1.5 over the whole breast volume (mean ± standard deviation). Computer simulations predict a reduction in the dynamic range if field heterogeneity leads to unintended water suppression, and distortion to CA uptake curves due to fat suppression failure (for fat containing pixels). A compromise between dynamic range and fat saturation performance is required. Both water suppression and fat suppression failure are apparent in clinical examinations.
Conclusion:
Magnetic field heterogeneity is likely to reduce the sensitivity of DCE-FS by distorting the CA uptake curves because of fat suppression failure (for fat containing pixels) and by reducing the dynamic range because of unintended water suppression.
Key Points:
• Magnetic field heterogeneity is significant in breast magnetic resonance. • Contrast-agent uptake curves are distorted by a non-uniform magnetic field. • Radiologist must be aware of possibility of distortion to interpret uptake curves correctly. • Compromise between fat suppression and dynamic range is required.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI