Related Experiment Video
Updated: May 15, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
A fractured and immovable ceramic liner in a screw-fixed acetabular shell during total hip arthroplasty: a case
1Department of Orthopaedic Surgery, Royal Lancaster Infirmary, Lancaster, UK.
Abstract:
We report a complication during total hip arthroplasty using an uncemented press-fit acetabular shell and ceramic liner for a woman with osteoarthritis. Fracture of the ceramic liner occurred during its insertion owing to non-concentric reduction. The main body of the ceramic liner became wedged in the acetabular shell and was resistant to removal. The shell itself was immovable owing to the prior insertion of 2 polyaxial screws, to which access was covered. Successful removal involved drilling a hole in the ceramic liner as a stress riser, and then cracking the liner with a Steinman pin.

