Related Experiment Video
Updated: May 15, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Nasal high-frequency oscillation for lung carbon dioxide clearance in the newborn
Amit Mukerji1, Michael Finelli, Jaques Belik
1Division of Neonatology, Department of Pediatrics, The Hospital for Sick Children, University of Toronto, Toronto, Ont., Canada.
Background:
Noninvasive ventilation has been used increasingly in recent years to minimize the duration of endotracheal mechanical ventilation in neonates due to its association with lung injury. Nasal high-frequency oscillation (nHFO) is a relatively new noninvasive modality but evidence for its use is limited.
Objective:
The goal of this study was to compare the CO2 clearance efficacy of nHFO and noninvasive positive pressure ventilation (NIPPV) in a neonatal lung model.
Design/Methods:
A newborn mannequin with dimensions and anatomy similar to a term infant was utilized. It was connected to a commercially available neonatal mechanical ventilator using a manufacturer-provided nasal adaptor. Various modes of noninvasive ventilation were compared as CO2 clearance was measured at the oropharynx by an end-tidal CO2 analyzer following the addition of a known amount of CO2 into the lung. Measurements were obtained at two different lung compliances using nHFO and compared with nCMV and nasal continuous positive airway pressure (nCPAP) as a control. Pressures near the nasal adaptor and the larynx were simultaneously measured with in-line pressure transducers.
Results:
Whereas no CO2 elimination was observed under nCPAP, its clearance with nHFO was 3-fold greater as compared to NIPPV. On nHFO, CO2 clearance was inversely proportional to frequency and maximal at 6 and 8 Hz. At a lower lung compliance, CO2 clearance was significantly higher at 6 Hz as compared to 10 Hz. During nHFO set to deliver a MAP of 10.0, we documented pressures of 7.2 ± 0.3 at the nasal adaptor and only 2.3 ± 0.3 cm H2O at the larynx.
Conclusions:
Nasal HFO is effective and superior to NIPPV at lung CO2 elimination in a newborn mannequin model. The use of nHFO as the preferred mode of noninvasive ventilation warrants further clinical studies.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Suctioning the Nasopharyngeal Airway
Equipment Required
Pulmonary Cycle: Exhalation
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...