ECG-based signal analysis technologies for evaluating patients with acute coronary syndrome: a systematic review
Philip J Leisy1, Remy R Coeytaux, Galen S Wagner
1Duke Evidence-based Practice Center, Duke Clinical Research Institute, Duke University, Durham, NC, USA.
Background/Objectives:
Timely identification of cardiac ischemia is critical in patients with acute coronary syndrome (ACS). The first test is often the standard, resting 12-lead ECG. Given its limitations, signal analysis enhancements have been proposed. We summarize the published evidence for commercially available ECG-based signal analysis technologies.
Methods:
This is a systematic review of the English-language published literature.
Results:
Published evidence meeting inclusion criteria was available for two devices: PRIME ECG and LP 3000. Meta-analysis of eight studies estimated a 68.4% sensitivity (95% CI, 35.1%-89.7%) and 91.4% specificity (CI, 83.6%-95.7%) for the PRIME ECG, compared with 40.5% sensitivity (CI, 19.6%-65.5%) and 95.0% specificity (CI, 87.9%-98.0%) for the standard 12-lead ECG.
Conclusions:
Existing evidence is insufficient to confidently inform the appropriate use of ECG-based signal analysis technologies for detecting ischemia or infarct in ACS. Further research is needed to determine in what circumstances, if any, these devices might precede, replace, or add to the standard ECG in test strategies for detecting ischemia or infarct in ACS.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Acute Coronary Syndrome I: Introduction
Dysrhythmias V: Evaluating Dysrhythmias
Acute Coronary Syndrome IV: Interprofessional Care