Development and validation of a reversed phase liquid chromatographic method for analysis of oxytetracycline and
Getu Kahsay1, Fairouz Shraim, Philippe Villatte
1Laboratorium voor Farmaceutische Analyse, Faculteit Farmaceutische Wetenschappen, Katholieke Universiteit Leuven, O&N2, PB-923, Herestraat 49, B-3000 Leuven, Belgium.
Abstract:
A simple, robust and fast high-performance liquid chromatographic method is described for the analysis of oxytetracycline and its related impurities. The principal peak and impurities are all baseline separated in 20 min using an Inertsil C₈ (150 mm × 4.6 mm, 5 μm) column kept at 50 °C. The mobile phase consists of a gradient mixture of mobile phases A (0.05% trifluoroacetic acid in water) and B (acetonitrile-methanol-tetrahydrofuran, 80:15:5, v/v/v) pumped at a flow rate of 1.3 ml/min. UV detection was performed at 254 nm. The developed method was validated for its robustness, sensitivity, precision and linearity in the range from limit of quantification (LOQ) to 120%. The limits of detection (LOD) and LOQ were found to be 0.08 μg/ml and 0.32 μg/ml, respectively. This method allows the separation of oxytetracycline from all known and 5 unknown impurities, which is better than previously reported in the literature. Moreover, the simple mobile phase composition devoid of non-volatile buffers made the method suitable to interface with mass spectrometry for further characterization of unknown impurities. The developed method has been applied for determination of related substances in oxytetracycline bulk samples available from four manufacturers. The validation results demonstrate that the method is reliable for quantification of oxytetracycline and its impurities.
More Related Videos
05:46Identification of Pharmaceuticals in The Aquatic Environment Using HPLC-ESI-Q-TOF-MS and Elimination of Erythromycin Through Photo-Induced Degradation
Published on: August 1, 2018
06:14Estimating the Yield of Compounds on the TLC Plate via the Blue-LED Illumination Technique
Published on: October 6, 2022
Related Concept Videos
Ion-Exchange Chromatography
Thin-Layer Chromatography (TLC): Overview
To begin the analysis, a mixture of compounds is spotted on the starting line on the TLC plate using a thin capillary. The bottom of the...