K Andersson1, I R Manchester, K Laurell
1Department of Radiation Sciences, Biomedical Engineering, Umeå University, Umeå, Sweden. kennet.andersson@radfys.umu.se
Introduction:
Infusion tests are used to diagnose and select patients with idiopathic normal pressure hydrocephalus (INPH) for shunt surgery. The test characterizes cerebrospinal fluid dynamics and estimates parameters of the cerebrospinal fluid system, the pressure-volume index (PVI) and the outflow conductance (Cout). The Oscillating Pressure Infusion (OPI) method was developed to improve the test and reduce the investigation time. The aim of this study was to evaluate the new OPI method by comparing it with an established reference method.
Methods:
Forty-seven patients (age 71.2 ± 8.9 years) with communicating hydrocephalus underwent a preoperative lumbar infusion investigation with two consecutive infusion protocols, reference (42 min) and new (20 min), that is, 94 infusion tests in total. The OPI method estimated Cout and PVI simultaneously. A real-time analysis of reliability was applied to investigate the possibility of infusion time reduction.
Results:
The difference in Cout between the methods was 1.2 ± 1.8 μl/s/kPa (ΔRout = -0.8 ± 3.5 mmHg/ml/min), P < 0.05, n = 47. With the reliability analysis, the preset 20 min of active infusion could have been even further reduced for 19 patients to between 10 and 19 min. PVI was estimated to 16.1 ± 6.9 ml, n = 47.
Conclusions:
The novel Oscillating Pressure Infusion method produced real-time estimates of Cout including estimates of reliability that was in good agreement with the reference method and allows for a reduced and individualized investigation time.
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Cerebrospinal Fluid
CSF Production
CSF is produced mainly in the choroid plexus, a network of capillaries and ependymal cells located within the ventricular system of the brain.