Evaluation of alternatives for dysfunctional double lumen central venous catheters using a two-compartmental
Koen Van Canneyt1, Wim Van Biesen, Raymond Vanholder
1Institute Biomedical Technology IBiTech-bioMMeda, Ghent University, Ghent, Belgium.
Abstract:
Double lumen (DL) central venous catheters (CVC) often suffer from thrombosis, fibrin sheet formation, and/or suction towards the vessel wall, resulting in insufficient blood flow during hemodialysis. Reversing the catheter connection often restores blood flows, but will lead to higher recirculation. Single lumen (SL) CVCs have often fewer flow problems, but they inherently have some degree of recirculation. To assist bedside clinical decision making on optimal catheter application, we investigated mathematically the differences in dialysis adequacy using different modes of access with CVCs. A mathematical model was developed to calculate reduction ratio (RR) and total solute removal (TSR) of urea, methylguanidine (MG), beta-2-microglobulin (β2M), and phosphate (P) during different dialysis scenarios: 4-h dialysis with a well-functioning DL CVC (DL-normal, blood flow QB 350 ml/min), dysfunctional DL CVC (DL-low flow, QB 250), reversed DL CVC (DL-reversed, QB 350, recirculation R = 10%) and 12 Fr SL CVC (effective QB 273). With DL-normal as reference, urea RR was decreased by 3.5% (DL-reversed), 13.0% (SL), and 15.6% (DL-low flow), while urea TSR was decreased by 3.3% (DL-reversed), 13.2% (SL), and 13.5% (DL-low flow). The same trend was found for MG and P. However, β2M RR decreased only 1.5% with SL CVC although TSR decrease was 17.2%, while RR decreased 21.1% with DL-low flow although TSR decrease was only 4.9%. In the case of dysfunctional DL CVCs, reversing the catheter connection and restoring the blood flow did not impair TSR, with 10% recirculation. The SL CVC showed suboptimal TSR results that were similar to those of the dysfunctional DL CVC.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
09:17Use of a Central Venous Line for Fluids, Drugs and Nutrient Administration in a Mouse Model of Critical Illness
Published on: May 2, 2017
Related Concept Videos
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Two-Compartment Open Model: Extravascular Administration
The absorption exponent (ka) indicates the speed at which the drug is...
Two-Compartment Open Model: Overview
The...
Three-Compartment Open Model
Compartment Models: Two-Compartment Model
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...