[Comparison between 226 Hz probe tone tympanometry and spiral CT test in infants]

Zhiqi Liu1, Lisi Liu, Kun Yang

  • 1Department of Otolaryngology Head and Neck Surgery, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430030, China.

Insights

226 Hz tympanometry is not recommended for diagnosing middle ear fluid in infants under 7 months. This hearing test showed low diagnostic value when compared to CT scans in the study.

Area of Science:

  • Pediatric Audiology
  • Otolaryngology

Background:

  • Middle ear effusion (MEE) is common in infants, potentially impacting hearing.
  • Accurate diagnosis of MEE is crucial for timely intervention.

Purpose of the Study:

  • To assess the diagnostic accuracy of 226 Hz probe-tone tympanometry for detecting middle ear fluid in infants aged 3-7 months.

Main Methods:

  • Fifty-two infants underwent 226 Hz tympanometry and spiral computerized tomography (CT) scans.
  • Tympanometric results were compared with CT findings to determine concordance.
  • Statistical analysis was performed using SPSS 16.0.

Main Results:

  • CT identified MEE in 57 out of 104 ears.
  • The concordance rate between 226 Hz tympanometry and CT for diagnosing MEE was low (kappa=0.103).
  • Specific tympanogram types (A, B, C, D) showed varied distributions in both normal and effusion ears.

Conclusions:

  • 226 Hz tympanometry demonstrates limited diagnostic value for identifying middle ear fluid in infants under 7 months.
  • Clinical use of 226 Hz tympanometry for MEE detection in this age group is not advised.
Abstract