A validated high-performance liquid chromatography method with diode array detection for simultaneous determination
Tian-Zeng Niu1, Yu-Wei Zhang, Yong-Li Bao
1National Engineering Laboratory for Durggable Gene and Protein Screening, Northeast Normal University, Changchun 130024, China.
Abstract:
A reversed phase high performance liquid chromatography method coupled with a diode array detector (HPLC-DAD) was developed for the first time for the simultaneous determination of 9 flavonoids in Senecio cannabifolius, a traditional Chinese medicinal herb. Agilent Zorbax SB-C18 column was used at room temperature and the mobile phase was a mixture of acetonitrile and 0.5% formic acid (v/v) in water in the gradient elution mode at a flow-rate of 1.0mlmin(-1), detected at 360nm. Validation of this method was performed to verify the linearity, precision, limits of detection and quantification, intra- and inter-day variabilities, reproducibility and recovery. The calibration curves showed good linearities (R(2)>0.9995) within the test ranges. The relative standard deviation (RSD) of the method was less than 3.0% for intra- and inter-day assays. The samples were stable for at least 96h, and the average recoveries were between 90.6% and 102.5%. High sensitivity was demonstrated with detection limits of 0.028-0.085μg/ml for flavonoids. The newly established HPLC method represents a powerful technique for the quality assurance of S. cannabifolius.