Troubleshooting carry-over of LC-MS/MS method for rifampicin, clarithromycin and metabolites in human plasma
D H Vu1, R A Koster, A M A Wessels
1Department of Pharmacotherapy and Pharmaceutical Care, University of Groningen, Groningen, The Netherlands.
Abstract:
Clarithromycin and rifampicin are used for the treatment of Mycobacteria. Pharmacokinetic drug interaction is possibly due to the influence of the two drugs on the liver enzymes. Using a Hypurity Aquastar C18 column (50mm×2.1mm×5μm) for liquid chromatography including a polar end-capped phase for the determination of clarithromycin, rifampicin and their metabolites together in plasma using LC-MS/MS resulted in a substantial carry-over. As a consequence, the throughput of the method is not assured. Using a step-by-step troubleshooting procedure, such carry-over was found originating from column memory effect. With the use of another type of C18 column, the carry-over is eliminated. Due to the absence of carry-over, the analytical concentration ranges are extended and are therefore more appropriate for the analysis of patient samples. The method was re-validated for linearity, reproducibility and dilution integrity.