Variance of SUVs for FDG-PET/CT is greater in clinical practice than under ideal study settings
Virendra Kumar1, Kavindra Nath, Claudia G Berman
1Department of Imaging, H Lee Moffitt Cancer Center and Research Institute, Tampa, FL 33612-9416, USA.
Purpose:
Measurement variance affects the clinical effectiveness of PET-based measurement as a semiquantitative imaging biomarker for cancer response in individual patients and for planning clinical trials. In this study, we measured test-retest reproducibility of SUV measurements under clinical practice conditions and recorded recognized deviations from protocol compliance.
Methods:
Instrument performance calibration, display, and analyses conformed to manufacture recommendations. Baseline clinical (18)F-FDG PET/CT examinations were performed and then repeated at 1 to 7 days. Intended scan initiation uptake period was to repeat the examinations at the same time for each study after injection of 12 mCi FDG tracer. Avidity of uptake was measured in 62 tumors in 21 patients as SUV for maximum voxel (SUV(max)) and for a mean of sampled tumor voxels (SUV(mean)).
Results:
The range of SUV(max) and SUV(mean) was 1.07 to 21.47 and 0.91 to 14.69, respectively. Intraclass correlation coefficient between log of SUV(max) and log of SUV(mean) was 0.93 (95% confidence interval [CI], 0.88-0.95) and 0.92 (95% CI, 0.87-0.95), respectively.Correlation analysis failed to show an effect on uptake period variation on SUV measurements between the 2 examinations, suggesting additional sources of noise.The threshold criteria for relative difference from baseline for the 95% CI were ± 49% or ± 44% for SUV(max) or SUV(mean), respectively.
Conclusions:
Variance of SUV for FDG-PET/CT in current clinical practice in a single institution was greater than expected when compared with benchmarks reported under stringent efficacy study settings. Under comparable clinical practice conditions, interpretation of changes in tumor avidity in individuals and assumptions in planning clinical trials may be affected.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
10:04Analysis of 18FDG PET/CT Imaging as a Tool for Studying Mycobacterium tuberculosis Infection and Treatment in Non-human Primates
Published on: September 5, 2017
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
![Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography (FDG-PET/CT)](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F3777.jpg&w=3840&q=50)