Related Experiment Videos
A balloon-expandable intravascular stent for obliterating experimental aortic dissection
M S Trent1, V Parsonnet, R Shoenfeld
1Division of Vascular Surgery, Newark Beth Israel Medical Center, NJ 07112.
Journal of Vascular Surgery
|May 1, 1990
Summary
A novel stent effectively obliterates acute aortic dissections in dogs, preventing rupture and maintaining branch vessel patency. This innovative treatment shows promise for managing this life-threatening condition.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Vascular Biology
Background:
- Acute aortic dissection is a critical condition requiring immediate intervention.
- Current treatments include aggressive hypotensive therapy and emergent surgery.
- A new intravascular stent offers a potential solution for internal obliteration of aortic dissections.
Purpose of the Study:
- To evaluate the efficacy of a balloon-expandable intravascular stent in treating experimental acute aortic dissections.
- To assess the stent's safety, including migration, aortic integrity, and branch vessel patency.
- To examine the histological response to stent implantation in the aorta.
Main Methods:
- A flexible, cut-to-length stent was implanted in the thoracic and abdominal aortas of 12 dogs (Group I).
- Thoracic aortic dissections were surgically created in 24 dogs; 12 received immediate stent implantation (Group II), and 12 were observed for natural history (Group III).
- Angiography and microscopy were performed post-implantation and post-explantation to assess outcomes.
Main Results:
- In Group I, no stent migration or aortic rupture occurred; all branch vessels remained patent with neointimal incorporation.
- In Group II, all dissections were obliterated by the stent.
- Group III showed high mortality (25%) and persistence of dissection (67%) without stenting, while stenting in persistent dissections led to obliteration.
Conclusions:
- The tested stent effectively obliterates experimental acute aortic dissections.
- The stent demonstrated safety, maintaining aortic integrity and branch vessel patency without thrombosis.
- This intravascular stent shows significant potential for treating acute aortic dissections.