A precise charge balancing and compliance voltage monitoring stimulator front-end for 1024-electrodes retinal
Hosung Chun1, Nhan Tran, Yuanyuan Yang
1Department of Electrical and Electronic Engineering, University of Melbourne, Melbourne, Australia. hosung.chun@unimelb.edu.au
Abstract:
In this paper, we present a precise charge balancing and compliance voltage monitoring stimulator front-end for 1024-electrode retinal prosthesis. Our stimulator is based on current mode stimulation. To generate a precisely matched biphasic current pulse, a dynamic current copying technique is applied at the stimulator front-end. A compliance voltage monitoring circuitry is included at the stimulator front-end to detect if a voltage across electrode-tissue interface goes beyond a predefined compliance voltage. Simulation results show the mismatch of a biphasic current pulse (at a maximum stimulation current of 476µA) is less than 0.1%. Also, the stimulator issues alarm signals, when a voltage compliance occurs during stimulation due to high tissue impedance. Our stimulator is implemented using a 65nm low voltage (LV) CMOS process, which helps reducing implementation area and power consumption.