Accuracy of end-tidal CO2 capnometers in post-cardiac surgery patients during controlled mechanical ventilation
Serge J H Heines1, Ulrich Strauch, Paul M H J Roekaerts
1Department of Intensive Care, Maastricht University Medical Centre, Maastricht, The Netherlands.
Background:
The determination of end-tidal carbon dioxide (etCO2) is very helpful in cardiac resuscitation for confirmation and monitoring of endotracheal tube placement and as an indicator of return of circulation and effectiveness of chest compressions. There is now also widespread use of capnometry on-site at emergency and trauma fields.
Objective:
We studied the accuracy and correlation of three capnometers (EMMA, Medtronic, and Evita) with partial pressure of arterial CO2 (PaCO2) measurements.
Methods:
The three capnometers were placed in-line in the ventilator tubing of the patient. Forty sedated and mechanically ventilated post-cardiac surgery patients were studied. Twenty consecutive etCO2 values were collected simultaneously from all three monitors while drawing an arterial blood sample. Paired sample t-test and Pearson correlation were used to compare the capnometers and their correlation with PaCO2.
Results:
The correlation of etCO2 measurements between all three capnometers was good (Emma vs. Evita: 0.874, Emma vs. Medtronic: 0.949, Evita vs. Medtronic: 0.878). The correlation of PaCO2 with the Evita is the lowest (0.671) as compared to the EMMA (0.693) and the Medtronic (0.727). The lowest dispersion of the difference between etCO2 and PaCO2 was seen in EMMA (3.30), the highest in Evita (3.98).
Conclusions:
A good correlation between etCO2 and PaCO2 was shown with the three capnometers in the present study. However, etCO2 measurements were not valid to estimate PaCO2 in these patients. Therefore, capnometry cannot be used to replace serial blood gas analyses completely, but may be a good cardiopulmonary trend monitor and alerting system in catastrophic events.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Acute Respiratory Failure-III
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiomyopathy VII: Pre and Post Operative Nursing Management