Noninvasive photoacoustic detecting intraocular foreign bodies with an annular transducer array
Diwu Yang1, Lvming Zeng, Changning Pan
1Department of Physics and Engineering, Hunan University of Technology, Zhuzhou 412000, China. yangdiwu@163.com
Abstract:
We present a fast photoacoustic imaging system based on an annular transducer array for detection of intraocular foreign bodies. An eight-channel data acquisition system is applied to capture the photoacoustic signals using multiplexing and the total time of data acquisition and transferring is within 3 s. A limited-view filtered back projection algorithm is used to reconstruct the photoacoustic images. Experimental models of intraocular metal and glass foreign bodies were constructed on ex vivo pig's eyes and clear photoacoustic images of intraocular foreign bodies were obtained. Experimental results demonstrate the photoacoustic imaging system holds the potential for in clinic detecting the intraocular foreign bodies.