Achieving targeted pulse oximetry values in preterm infants in the delivery room
Bheru Gandhi1, Wade Rich, Neil Finer
1Division of Neonatology, University of California San Diego Medical Center, San Diego, CA 92103, USA.
Objective:
To determine whether resuscitation teams can better maintain newborn transitional oxygen saturation (SpO₂) values within a prespecified target range using a graphical display of the targets and real-time SpO₂ data compared with using only numerical oximeter values.
Study Design:
Preterm neonates were enrolled in this prospective cohort evaluation of a change in practice. The Transitional Oxygen Targeting System (TOTS) plots real-time SpO₂ values in relation to 10th and 50th percentile SpO₂ curves, which provides a visual target. After introduction of the TOTS, the resuscitation team adjusted the fraction of inspired oxygen to maintain the SpO₂ within the target range, and before its use neonates were resuscitated in attempt to target normal transitional SpO₂ values without using the TOTS monitor. Duration of time within 10th-50th percentile values was compared between 2 cohorts, children evaluated with the TOTS and those not evaluated with the TOTS (controls).
Results:
A total of 40 infants were enrolled, including 20 prospectively evaluated with the TOTS and 20 controls. Mean gestational age, birth weight, total resuscitation time, duration of supplemental oxygen administration, changes in oxygen concentration, and respiratory support provided were similar in the 2 groups. The TOTS cohort spent 52% of time within the target range; the control cohort, 37% (P = .03).
Conclusion:
SpO₂ values were maintained within a specified target range for significantly longer in preterm neonates resuscitated using the TOTS display compared with those resuscitated without TOTS.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...