Ex Vivo OCT-Based Multimodal Imaging of Human Donor Eyes for Research into Age-Related Macular Degeneration
Published on: May 26, 2023
Wide-angle fundus imaging through the Boston keratoprosthesis
Rony R Sayegh1, Claes H Dohlman
1Department of Ophthalmology, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
Retina (Philadelphia, Pa.)
|February 19, 2013
Summary
Wide-angle fundus cameras can image the peripheral retina through a Boston keratoprosthesis. The noncontact Optos camera offers superior image quality and ease of use compared to contact devices.
Area of Science:
- Ophthalmology
- Medical Imaging
Background:
- Keratoprosthesis implantation is a surgical procedure to restore vision in patients with severe corneal disease.
- Imaging the peripheral retina in these patients presents unique challenges due to the artificial cornea.
Purpose of the Study:
- To evaluate the feasibility of using three wide-angle fundus cameras for peripheral retinal imaging in eyes with a Type 1 Boston keratoprosthesis.
- To compare the image quality, ease of acquisition, and limitations of noncontact (Optos) and contact (RetCam, Panoret) imaging systems.
Main Methods:
- Noncontact Optos and contact RetCam and Panoret wide-angle imaging systems were used to image the retina in eyes with keratoprosthesis.
- Failure-to-image rates, image quality, field of view, and complications were assessed.
- Optos imaging findings were correlated with ultrasound B-scan results.
Main Results:
- Wide-angle retinal images were successfully obtained with all three cameras in some eyes.
- Optos achieved the highest success rate (all four eyes), followed by RetCam (three eyes) and Panoret (two eyes).
- Optos provided the best image quality and was preferred for its noncontact nature, ease of use, and safety, outperforming B-scan in specific cases.
Conclusions:
- Wide-angle fundus imaging through a keratoprosthesis is feasible with available technologies.
- The noncontact Optos system is a preferred choice due to its superior image quality, ease of use, and patient comfort.
- Optos imaging complements and may surpass ultrasound B-scan for peripheral retinal examination in keratoprosthesis patients.