Related Experiment Video
Updated: May 13, 2026

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Loss and retrieval of a coronary angioplasty stent balloon
Peter Kayaert1, Jeroen Sonck, Oscar Semeraro
1Universitair Ziekenhuis (UZ) Brussel, Laarbeeklaan 101, 1090 Brussel, Belgium. peter.kayaert@uzbrussel.be
Abstract:
A case is presented in which a breakage of the coronary stent delivery catheter occurred as retrieval of the stent balloon was attempted after stent implantation. The broken distal balloon shaft with its stuck balloon was retrieved by controlled pulling on the guiding catheter in which a new balloon was inflated, thus trapping the distal shaft. The case underscores the importance of adequate lesion assessment and preparation.
Related Concept Videos
Aneurysm III: Interprofessional Care
Coronary Artery Disease V: Interprofessional Care
Aneurysm I: Introduction
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization