Comparison of anaesthesia ventilators using a lung model.
1Department of Anesthesiology, University Hospital, Linköping University, Sweden.
Acta Anaesthesiologica Scandinavica
|April 1, 1990
Summary
Six anesthesia ventilators were tested for ventilatory performance. The Servo Ventilator 900 D demonstrated superior gas distribution compared to other models and the Engström 300.
Area of Science:
- Anesthesiology
- Mechanical Ventilation
- Respiratory Mechanics
Background:
- Evaluating the performance of anesthesia ventilators is crucial for patient safety.
- Older models like the Engström 300 set a benchmark for ventilatory performance.
- Advancements in ventilator technology necessitate comparative performance assessments.
Purpose of the Study:
- To compare the ventilatory performance of six modern anesthesia ventilators against the established Engström 300.
- To assess the accuracy of delivered minute volume across different ventilators.
- To evaluate the effectiveness of gas distribution in simulated obstructed lungs.
Main Methods:
- Utilized a lung model with adjustable bronchial occlusion and simulated lungs.
- Compared dialled (set) minute volume with recorded minute volume for each ventilator.
- Calculated gas distribution between obstructed and non-obstructed lung compartments.
Main Results:
- All tested ventilators exhibited a linear relationship between dialled and measured minute volume.
- The Servo Ventilator 900 D showed gas distribution performance equal to or better than the Engström 300.
- Other ventilators' gas distribution capabilities varied, with some not matching the benchmark.
Conclusions:
- The Servo Ventilator 900 D is a high-performing anesthesia ventilator regarding gas distribution.
- Modern anesthesia ventilators show consistent minute volume delivery but varying gas distribution.
- Further research may be needed to optimize gas distribution in ventilators for complex respiratory scenarios.