A clinical evaluation of the Maquet Quadrox-i Neonatal oxygenator with integrated arterial filter
R M Ginther1, R Gorney, R Cruz
1Division of Pediatric Cardiothoracic Surgery, University of Texas Southwestern Medical Center and Children's Medical Center, Dallas, Texas 75235, USA. Richard.ginther@utsouthwestern.edu
Abstract:
High-performance, low-prime-volume oxygenators for the pediatric patient population have become a growing market among manufacturers. In the summer of 2011, our institution clinically evaluated the performance of the newly released Maquet Quadrox-i Neonatal oxygenator with integrated arterial filter. The static priming volume, including the integrated arterial filter, is 40 ml and the maximum rated blood flow is 1.5 liters per minute (LPM). The device was used on seven pediatric patients, ranging from 3.2 to 14 kg, undergoing various congenital heart defect repairs. Data were collected to calculate gas transfer, trans-oxygenator pressure drop, and heat exchange performance. The mean cardiopulmonary bypass time was 85 minutes and the mean cross-clamp time was 56 minutes. The average oxygen transfer was 34.3 ± 22.8 ml/O2/min and increased with both blood flow and FiO2. The average carbon dioxide transfer was 22.3 ± 17.8 ml/min and increased with both blood flow and gas sweep to blood flow ratio. The average trans-oxygenator pressure drop per blood flow was 53.3 ± 15.5 mmHg/L/min and increased with flow. The average heat exchanger performance factor was 47.6 ± 11.6% and decreased with flow. The heat exchange performance factor at maximum observed clinical flow, 1.42 LPM, was 36.4%. During this evaluation, the Maquet Quadrox-i Neonatal oxygenator adequately performed within its operational flow in the clinical setting.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: