Highly accelerated T1-weighted abdominal imaging using 2-dimensional controlled aliasing in parallel imaging results
Philipp Riffel1, Ulrike I Attenberger, Stephan Kannengiesser
1Department of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, 68167 Mannheim, Germany. Philipp.Riffel@umm.de
Purpose:
The purpose of this study was to evaluate the feasibility and technical quality of an abdominal 3-dimensional interpolated breath-hold (volumetric interpolated breath-hold examination [VIBE]) magnetic resonance examination using the new parallel acquisition technique, controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA).
Materials And Methods:
In this institutional review board-approved study, 15 volunteers underwent an abdominal magnetic resonance imaging examination including axial unenhanced 3-dimensional VIBE sequences with the conventional parallel acquisition technique, generalized autocalibrating partially parallel acquisitions parallel imaging (GRAPPA), with an acceleration factor (R) of 2, 3, 4, and 2 × 2 in comparison with a CAIPIRINHA-VIBE sequence with an acceleration factor of 2 × 2. Images were evaluated regarding the overall image quality, liver edge sharpness, and parallel imaging artifacts. Signal-to-noise ratio was evaluated using 2 different methods. In a second study population, 17 patients were examined with our new routine protocol for abdominal imaging that now comprises VIBE sequences with CAIPIRINHA with R = 2 × 2.
Results:
In the volunteer population, the overall image quality of CAIPIRINHA with R = 2 × 2 was significantly higher compared with GRAPPA with R = 3, 4, and 2 × 2 (P < 0.05). There were significantly less parallel imaging artifacts with CAIPIRINHA with R = 2 × 2 (P < 0.05). Acquisition time varied between 21.1 (GRAPPA with R = 2, 320 matrix) and 6.9 seconds (CAIPIRINHA with R = 2 × 2, 256 matrix). Signal-to-noise ratio performance of CAIPIRINHA with R = 2 × 2 was superior to GRAPPA with R = 3, 4, and 2 × 2. In the patient population, VIBE sequences with CAIPIRINHA with R = 2 × 2 showed consistently good image quality, minimal motion artifacts, and minimal parallel imaging artifacts.
Conclusions:
The CAIPRINHA-VIBE with an acceleration factor of R = 2 × 2 is feasible in a clinical setting and is characterized by fast and robust imaging with an image quality comparable with a 2-fold acceleration with GRAPPA.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...