MRI and PET in Mouse Models of Myocardial Infarction
Published on: December 19, 2013
Cardiac motion compensation and resolution modeling in simultaneous PET-MR: a cardiac lesion detection study
1Center for Advanced Medical Imaging Sciences, Nuclear Medicine and Molecular Imaging, Department of Radiology, Massachusetts General Hospital, 55 Fruit Street, Boston, MA 02114, USA.
Abstract:
Cardiac motion and partial volume effects (PVE) are two of the main causes of image degradation in cardiac PET. Motion generates artifacts and blurring while PVE lead to erroneous myocardial activity measurements. Newly available simultaneous PET-MR scanners offer new possibilities in cardiac imaging as MRI can assess wall contractility while collecting PET perfusion data. In this perspective, we develop a list-mode iterative reconstruction framework incorporating both tagged-MR derived non-rigid myocardial wall motion and position dependent detector point spread function (PSF) directly into the PET system matrix. In this manner, our algorithm performs both motion 'deblurring' and PSF deconvolution while reconstructing images with all available PET counts. The proposed methods are evaluated in a beating non-rigid cardiac phantom whose hot myocardial compartment contains small transmural and non-transmural cold defects. In order to accelerate imaging time, we investigate collecting full and half k-space tagged MR data to obtain tagged volumes that are registered using non-rigid B-spline registration to yield wall motion information. Our experimental results show that tagged-MR based motion correction yielded an improvement in defect/myocardium contrast recovery of 34-206% as compared to motion uncorrected studies. Likewise, lesion detectability improved by respectively 115-136% and 62-235% with MR-based motion compensation as compared to gating and no motion correction and made it possible to distinguish non-transmural from transmural defects, which has clinical significance given the inherent limitations of current single modality imaging in identifying the amount of residual ischemia. The incorporation of PSF modeling within the framework of MR-based motion compensation significantly improved defect/myocardium contrast recovery (5.1-8.5%, p < 0.01) and defect detectability (39-56%, p < 0.01). No statistical difference was found in PET contrast and lesion detectability based on motion fields obtained with half and full k-space tagged data.
Insights
This study introduces a new PET-MRI reconstruction method to improve cardiac imaging quality. The technique corrects for cardiac motion and partial volume effects, significantly enhancing lesion detection and contrast recovery for better diagnosis.
Area of Science:
- Medical Imaging
- Nuclear Medicine
- Cardiovascular Imaging
Background:
- Cardiac PET imaging is degraded by motion and partial volume effects (PVE), leading to artifacts and inaccurate measurements.
- Simultaneous PET-MR scanners enable concurrent assessment of myocardial contractility and PET perfusion data.
- Existing methods struggle with accurate quantification and defect characterization in cardiac PET.
Purpose of the Study:
- To develop a list-mode iterative reconstruction framework for cardiac PET-MR that incorporates tagged-MR derived motion and detector point spread function (PSF) modeling.
- To evaluate the effectiveness of this framework in improving image quality, contrast recovery, and lesion detectability using a beating cardiac phantom.
- To assess the impact of MR data acquisition time (full vs. half k-space) on motion correction accuracy.
Main Methods:
- Developed a list-mode iterative reconstruction framework integrating non-rigid myocardial motion from tagged-MR and position-dependent detector PSF.
- Utilized non-rigid B-spline registration to derive wall motion information from tagged MR volumes acquired with full or half k-space data.
- Evaluated the method in a beating cardiac phantom with transmural and non-transmural cold defects.
Main Results:
- MR-based motion correction improved defect/myocardium contrast recovery by 34-206% and lesion detectability by 115-235% compared to uncorrected studies.
- The method enabled differentiation between transmural and non-transmural defects, crucial for assessing residual ischemia.
- Incorporating PSF modeling alongside MR-based motion compensation further enhanced contrast recovery (5.1-8.5%) and lesion detectability (39-56%).
- No significant difference in PET contrast or lesion detectability was observed between full and half k-space tagged MR data.
Conclusions:
- The proposed PET-MR reconstruction framework effectively corrects for cardiac motion and PVE, significantly improving image quality and diagnostic accuracy.
- This integrated approach enhances the ability to detect and characterize myocardial defects, offering clinical advantages over current single-modality imaging.
- Acquisition of half k-space tagged MR data is sufficient for generating accurate motion fields, potentially accelerating imaging time without compromising results.