Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator.
James L Bedford1, Michael D R Thomas, Gregory Smyth
1Physics Department, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, Sutton, Surrey SM2 5PT, UK. James.Bedford@icr.ac.uk
Journal of Applied Clinical Medical Physics
|March 9, 2013
Summary
The Agility multileaf collimator (MLC) enables faster and accurate radiotherapy delivery. This study confirms Pinnacle3 accurately models the Agility MLC for volumetric-modulated arc therapy (VMAT), significantly reducing treatment times.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- The Agility multileaf collimator (MLC) offers advanced features for radiotherapy.
- Volumetric-modulated arc therapy (VMAT) requires precise beam delivery.
- Accurate modeling of MLCs in treatment planning systems is crucial for VMAT.
Purpose of the Study:
- To create and validate a Pinnacle3 treatment planning system beam model for the Agility MLC.
- To evaluate the performance of the Agility MLC in delivering VMAT treatments.
- To compare Agility MLC VMAT delivery times and accuracy against an MLCi unit.
Main Methods:
- Developed a 6 MV beam model for the Agility MLC in Pinnacle3.
- Verified the beam model using dose measurements under irregular fields.
- Created and delivered VMAT treatment plans for head-and-neck patients using Agility MLC with and without continuously variable dose rate (CVDR), comparing to an MLCi unit.
Main Results:
- The Agility MLC beam model demonstrated high accuracy, with interleaf leakage of 0.2% and mean dose agreement of 1.3% for irregular fields.
- Agility VMAT plans achieved equivalent quality and accuracy to MLCi, with 95% of measurements within 3% and 3 mm.
- Mean delivery times were significantly reduced: 133s (Agility with CVDR), 171s (Agility without CVDR), vs. 282s (MLCi).
Conclusions:
- Pinnacle3 accurately models the Agility MLC.
- Agility MLC facilitates accurate VMAT plan delivery.
- The Agility MLC significantly reduces VMAT treatment delivery time compared to MLCi.