Comparison of new and established full-field digital mammography systems in diagnostic performance
Eun Sook Ko1, Boo-Kyung Han, Sun Mi Kim
1Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul 135-710, Korea.
Objective:
To compare the diagnostic performance of new and established full-field digital mammography (FFDM) systems.
Materials And Methods:
During a 15-month period, 1038 asymptomatic women who visited for mammography were prospectively included from two institutions. For women with routine two-view mammograms from established FFDM systems, bilateral mediolateral oblique (MLO) mammograms were repeated using the new FFDM system. One of the four reviewers evaluated two-sets of bilateral MLO mammograms at 4-week intervals by using a five-point score for the probability of malignancy according to a Breast Imaging Reporting and Data System. The lesion type and breast density were determined by the consensus of two readers at each institution. The dichotomized mammographic results correlated with a final pathologic outcome and follow-up data. Receiver operating characteristic (ROC) curves, sensitivity, and specificity were compared in general and according to the lesion type and breast density.
Results:
Of the 1038 cases, 193 (18.6%) had cancer. The areas under the ROC curve (AUC), sensitivity, and specificity of the established system were 0.815, 65.3%, and 90.2%, respectively. Those of the new system were 0.839, 68.4%, and 91.7%, respectively. There were no significant differences in the AUCs, sensitivities or the specificities in general between new and established systems (Ps = 0.194, 0.590, 0.322, respectively). We found no significant difference in these parameters according to lesion type or breast density.
Conclusion:
The new FFDM system has a comparable diagnostic performance with established systems.