The use of video laryngoscopy during air medical transport: a manikin study
Joseph M Grover1, Jeffrey S Lubin, Angela L Finney
1Case Western Reserve University School of Medicine. jmg5dn@virginia.edu
Introduction:
This study simulated intubation with direct laryngoscopy and with a GlideScope Ranger video laryngoscope using a standard Laerdal airway manikin in a medical helicopter under various conditions. We hypothesized that the intubation times would be greater using direct laryngoscopy compared with the GlideScope under all conditions.
Methods:
Twenty crew members of a single helicopter emergency medical service participated in the study. Participants intubated an airway manikin using both direct laryngoscopy and the GlideScope Ranger in varying conditions, including standing in a room with the lights on and off, in the helicopter while stationary on the ground and unbelted during both daytime and nighttime, and finally in the aircraft while in flight belted during both daytime and nighttime. A study investigator recorded the intubation times and independently confirmed tracheal placement of the endotracheal tube.
Results:
For all 6 environments, the mean time for intubation was slightly greater using the GlideScope (18.7 seconds; 95% confidence interval, 17.4-20.0 seconds) compared with direct laryngoscopy (15.5 seconds; 95% confidence interval, 14.7-16.4). There was a statistically significant difference in times to intubation, in favor of direct laryngoscopy, in the settings of standing with the room lights on (P = 0.0013), on the ground in the helicopter unbelted during the daytime (P = 0.009), and in flight belted at nighttime (P = 0.0012), with the 3 other environments not reaching statistical significance.
Conclusions:
Using the GlideScope took more time to intubate compared with direct laryngoscopy in all tested environments. Although this difference in intubation times was statistically significant, it was not clinically significant, suggesting that both modalities may be comparable in nondifficult airways.
More Related Videos
04:56Laryngeal Mask Airway (LMA) Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway (SGA)
Published on: July 14, 2023
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...