Novel contrast-injection protocol for coronary computed tomographic angiography: contrast-injection protocol
Masafumi Kidoh1, Takeshi Nakaura, Shinichi Nakamura
1Diagnostic Radiology, Amakusa Medical Center, Kameba 854-1, Amakusa, Kumamoto, 863-0046, Japan, masafkidoh@yahoo.co.jp.
Abstract:
We developed a new individually customized contrast-injection protocol for coronary computed tomography (CT) angiography based on the time-attenuation response in a test bolus, and investigated its clinical applicability. We scanned 60 patients with suspected coronary diseases using a 64-detector CT scanner, who were randomly assigned to one of two protocols. In protocol 1 (P1), we estimated the contrast dose to yield a peak aortic attenuation of 400 HU based on the time-attenuation response to a small test-bolus injection (0.3 ml/kg body weight) delivered over 9 s. Then we administered a customized contrast dose over 9 s. In protocol 2 (P2), the dose was tailored to the patient's body weight; this group received 0.7 ml/kg body weight with an injection duration of 9 s. We compared the two protocols for dose of contrast medium, peak attenuation, variations in attenuation values of the ascending aorta, and the success rate of adequate attenuation (250-350 HU) of the coronary arteries. The contrast dose was significantly smaller in P1 than in P2 (36.9 ± 9.2 vs 43.1 ± 7.0 ml, P < 0.01). Peak aortic attenuation was significantly less under P1 than under P2 (384.1 ± 25.0 vs 413.5 ± 45.7, P < 0.01). The mean variation (standard deviation) of the attenuation values was smaller in P1 than in P2 (25.0 vs 45.7, P < 0.01). The success rate of adequate attenuation of the coronary arteries was significantly higher with P1 than with P2 (85.0 vs 65.8 %, P < 0.01). P1 facilitated a reduction in the contrast dose, reduced the individual variations in peak aortic attenuation, and achieved optimal coronary CT attenuation (250-350 HU) more frequently than P2.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
10:28Interventional Diagnostic Procedure: A Practical Guide for the Assessment of Coronary Vascular Function
Published on: March 15, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies VII: Vascular Imaging
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...